Validation and routine control of processes in the H2O2 (hydrogen peroxide) steriliser

During the validation and routine control of VH2O2 processes, target values specified by the manufacturer are measured, documented and assessed using independent measuring technology, such as pressure-temperature data loggers, in order to verify the sterilisation conditions. The new high-precision temperature and pressure logger from Xylem/ebro is ideal for recording the vacuum test and the sterilisation parameters of pressure, temperature and time. The data logger operates in a pressure measurement range from 0.1 to 1050 mbar (0.1 to 788 Torr) with an extremely high precision of +/- 0.25 mbar (0.1 mbar to 50 mbar measurement range), making it the perfect choice when measuring the vacuum in H2O2 sterilisers. The data logger also measures temperatures to within +/- 0.1 °C accuracy in a measurement range of 0 to 85 °C.

ebro. a xylem brand